Laboratory Incubators Capabilities
PHCbi brand laboratory incubators are designed to ensure highly accurate, reproducible environments that mimic real-world conditions. Our laboratory incubators allow for precise control of temperature, a low humidity setting and lighting protocols – making them ideal for microbiology testing and research, food and cosmetic shelf life testing, and municipal wastewater testing, plant cell culturing and more. Robust cabinet construction with corrosion-resistant stainless-steel surfaces simplify cleaning and eliminate rust due to high humidity in these heated and cooled incubators. PHCbi brand laboratory incubators use SNAP compliant, low flammability refrigerants to assure low environmental impact.
Versatile Applications
The PHCbi compact microbiological incubators offers accurate temperature control and consistent uniformity making them suitable for laboratory culturing, industrial incubation and various testing applications.
Wide Temperature Range
This plant cell biology incubator offers accurate temperature control and uniformity in wide temperature ranges (-10°C up to +60°C) with excellent chamber uniformity. The microprocessor allows for precision temperature control combined with a heater PID and compressor ON-OFF system. With a feed forward function that inputs the operating conditions of the compressor beforehand, the system ensures accurate temperature control for the chamber.
Ergonomic Cabinet Design
The MIR-154-PA features improved usability with a modern design, reversible door, adjustable low vibration and gentle air circulation that reduces media drying.
Microprocessor Controls
The MIR Series compact microbiological incubator has an 8-bit microprocessor controller for heat and refrigeration control +/-2°C. This compact microbiological incubator is completely programmable, allowing for direct numeric input with memory of operation, start dates and hours, clock mode and timer. The LCD control improves user interface for better programming and control.
Memory Backup
Should the power source be interrupted due to power failure or other event, programmed data remains stored in memory semi-permanently. When the power source is restored, operation can be continued according to the predetermined program.
Programmable Operational Functions
Program input is simple and accommodates a range of diversified experimentation requirements. It is ideal for night time or weekend experimentation or experimentation that requires setting changes. It is also ideal for microorganism cultures and preservation.
EFFECTIVE CAPACITY | 4.34 cu.ft. (123 L) |
---|---|
EXTERIOR DIMENSIONS | 27.6″ W x 22.8″ D x 40.1″ H (700 x 580 x 1018 mm) |
INTERIOR DIMENSIONS | 24.4″ W x 14.5″ D x 21.9″ H (620 x 368 x 555 mm) |
DOORS | Painted Steel |
INSULATION | Rigid polyurethane foamed-in-place |
SHELVES AND SHELF DIMENSIONS | Polyethylene coated steel wire, 3 pcs, adjustable. Dimensions: Upper 2 pcs, 22.4″ x 11.8″ (570 x 300 mm); Bottom 1 pc, 21.7″ x 9.25″ (550 x 235 mm); Max load: 44 lbs (20 kg) |
AIR CIRCULATION | Forced air circulation |
ACCESS PORT | 1: Left side (1.57″ (40 mm) diameter) |
COMPRESSOR | Reciprocated compressor. Output: 150 W |
TEMPERATURE CONTROL RANGE | -10ºC to 60ºC (ambient temperature: 5ºC to 35ºC, no load) |
TEMPERATURE DISPLAY | Microprocessor, PID Control (compressor; ON-OFF) |
TEMPERATURE UNIFORMITY | ±0.5ºC (SV:37%, ambient temperature: 20ºC, no load) |
ALARMS | High/Low temp alarm, independent over-heat/over-cool limiter |
NET WEIGHT | 172 lbs (78 kg) |
ELECTRICAL REQUIREMENTS | 160 W, AC 115 V, 60 Hz |
ENERGY CONSUMPTION | 200 W |
Reviews